IHCalibrators®

BCS IHCalibrators® make Calibration possible for IHC
  • 15 small round spots pre-printed on a microscope slide
  • The spots are glass microbeads covalently coated with a fluorescent biomarker
  • Each bead has a precise concentration of the biomarker traceable to a primary standard at NIST
  • The numbers provide an orientation to the sequence of increasing biomarker concentrations
  • A certificate of analysis provides the exact concentrations for each level

  • Identify the lowest concentration level that is still stained.
  • In each Panel, level 5 is the lowest that still has sufficient color intensity to consider positive.
  • Look at the Certificate of Analysis for that calibrator and correlate it with the actual concentration.
  • The actual Lower Limit of Detection (LOD) is likely at a concentration between levels 4 and 5.
  • Metric calculates the precise LOD.

  • IHCalibrators® enable this graph
  • Metric calculates LOD and dynamic range and establishes control ranges
  • Select an IHControl® within the dynamic range
  • Compare your LOD with other peer labs